7.5
Setting Screen
The LSP Series sensor measures the hydrostatic pressure created by the liquid level in the tank. Convert-
ing this measured pressure value to other parameters such as the tank level and the remaining liquid vol-
ume in gallons requires the tank volume capacity, the tank maximum liquid height measured from the
bottom of the tank to the liquid surface when filled to the rated capacity, and the density of the liquid.
Measurements assume the tank has a uniform horizontal cross-section. To convert the measured pressure
to volumetric information, the LSP Series sensor requires the user to enter three parameters via the
uPyxis®
Desktop App’s
Setting
screen:
•
Volume
(rated volume capacity of the tank)
•
Max Height
(from the bottom of the tank to the liquid surface when filled to the rated capacity)
•
Density
(of liquid in lb/gal)
From the
Settings
screen, you can also set the
Display Unit
and
Log Period
(time, in seconds, between two
measurements). To save your setting changes, click
Set
.
Figure 24.