44
The activation of the
Analog In
button allows to read in a voltage
signal at the extension port, e.g. from third-party sensors like a pH
sensor. The measuring range is 0-2.5V and the display and output
of the signal is in mV. For details refer to chapter 11.
The maximum number of data points kept in the graphs can be
changed by the selector
Max. Data Points in Graphs
(default:
10800). A change of the number will clear the graphs and high
values (>>10000) might decrease the maximum sample rate.
The
FireStingO2
offers four analog outputs (0-2.5V) at the
extension port which can be configured by pressing the
Analog
Output
button. For details refer to chapter 11.
Advanced Options
(only for advanced users)!
The
USB
communication
speed
can be adjusted e.g. for improving
the maximum sampling rate (default: 57600).
Activation of the button
Enable High-Speed Sampling
will enable
the adjustment of a Sample Interval <0.25 s in the main window
(and disable
Max. Data Points in Graph
).
The
Advanced Auto-Mode
button allows advanced configuration
options for the Auto-Mode. For details refer to chapter 11.