102
12.4
Measuring Principle
The new
REDFLASH technology
is based on the unique oxygen-
sensitive
REDFLASH indicator
showing excellent brightness. The
measuring principle is based on the quenching of the
REDFLASH
indicator
luminescence caused by collision between oxygen
molecules and the
REDFLASH indicator
immobilized on the sensor
tip or surface. The
REDFLASH indicators
are excitable with red
light (more precisely: orange-red at a wavelength of 610-630 nm)
and show an oxygen-dependent luminescence in the near infrared
(NIR, 760-790 nm).
The
REDFLASH technology
impresses by its high precision, high
reliability, low power consumption, low cross-sensitivity, and fast
response times. The red light excitation significantly reduces
interferences caused by autofluorescence and reduces stress in
biological systems. The
REDFLASH indicators
show much higher