Signal too high
There might be too much ambient light on the
sensor tip or the sensor spot. Avoid direct sun light exposure or
strong direct illumination with a lamp. (For advanced users:
decrease the LED intensity and/or the amplification in the
Advanced Settings)
Bad reference
This
indicates internal problems of the
FireStingO2
. Please contact
PyroScience
for support.
Refer also to Troubleshooting in chapter 12.3.
The adjustment of the Settings using the button
Settings
is
described in detail in chapter 6.2.
The button
Save Setup
can be used to save the current settings
and calibration data of all channels. They can be reloaded anytime
by pressing the button
Load Setup
. This allows e.g. to switch
between different laboratory setups with a single
FireStingO2
.
This function might be also useful if different computers are used
for the calibration and for the actual measurements. You might
calibrate the sensors with the first computer, save the
configuration with
Save Setup
. By transferring this file and also
the oxygen meter
FireStingO2
to a second computer, you can load
there again this configuration with
Load Setup
giving you
calibrated sensors ready for the measurement.
Save Setup
and
Load Setup
might be also useful if e.g. a 1-
channel
FireStingO2
should be used repeatedly with several
oxygen sensors. Initially each sensor needs to be calibrated only
once and the configuration of each sensor is saved with
Save
Setup
. If later on a measurement should be performed with a
specific sensor, it is only necessary to load the configuration for
this sensor with
Load Setup
.