might be an incorrect Sensor Code entered in the device or
software.
Too High: Avoid direct sun light exposure or strong direct
illumination with a lamp and darken the surrounding. And/or check
the entered Sensor Code.
In case of the warning Bad Ref, indicating internal problems of the
electronics, please contact
PyroScience
for support.
Bad Ref: Please do not continue to measure if a warning is shown!
11.2
Temperature Sensor Warning
If the
Temperature Sensor is activated in the
Settings (see chapters 6.1.4 and 10.3.2) for
automatic temperature compensation of the
oxygen measurements, but is not connected,
broken or nonfunctional, the warning No T
Sensor
is shown in the top screen of the Live Mode (device) and in
the Live Graph window of the
FireStingGO2 Manager
. While this
warning is active, the device cannot provide any
valid oxygen readings, as the compensation
temperature is missing for calculating the
oxygen value. The oxygen reading is then
replaced by "--.--".
No measurements can be
performed until the temperature sensor is
connected or exchanged. Alternatively, the
Settings can be changed to a defined Fixed
temperature (see chapters 6.2.4 and 10.3.2).
No T Sensor: Check the connection to the temperature sensor. Or
replace broken temperature sensor.
Also in the Logging Mode, a missing or non-functional
temperature sensor will lead to a disappearance of the oxygen and