47
10.3.6
Data Logging
Before starting data Logging, adjust the sensor Settings and
complete a sensor Calibration. To deploy the device after logging
has been started, please ensure that the
FireStingGO2 Manager
software is closed before the device is disconnected.
Even if the
FireStingGO2
is
operated with the
FireStingGO2
Manager
software, the logged
data are always saved within
the device memory of the
FireStingGO2
. After finalization
of the logging, the log file can
be downloaded from the device
to the PC. This feature ensures
that
the
device
can
be
disconnected without problems
from the PC, even when logging
is active. A typical scenario is
that the device is configured and calibrated with the
FireStingGO2
Manager
, then logging is started and the device is disconnected
and deployed at the measurement location. Later when the
measurements are completed, the device is again connected to the
PC, and the logged data are downloaded and inspected. The
following steps must be done for data logging:
(1) Device Time. Data logging is
managed in the Logging tab of the
main window. First check the current
device time (i.e. the time of the internal
clock in the
FireStingGO2
) shown in this
tab. The time and date information for
the logged data are always related to
this device time (and not the PC time).