Radiated RF   
IEC 61000-4-3 

3 V/m 
80 MHz to 2.5 GHz 

3V/m 
 

[EQUIPMENT  or  SYSTEM],  including 
cables, 

than 

the 

recommended 

separation  distance  calculated  from  the 
equation  applicable  to  the  frequency  or 
the transmitter. 
Recommended separation distance 

d = 1.2 

p

 

d = 1.2 

p

  80 MHz to 800 MHz 

d = 2.3 

p

  800 MHz to 2.5 GHz 

Where  P  is  the  maximum  output  power 
rating  of  the  transmitter  in  watts  (W) 
according 

to 

the 

transmitter 

manufacturer and d is the recommended 
separation distance in meters (m). 
Field 

strengths 

from 

fixed 

RF 

transmitters,  as  determined  by  an 
electromagnetic  site  survey,  should  be 
less  than  the  compliance  level  in  each 
frequency range. 
Interference may occur in the vicinity of 
equipment  marked  with  the  following 
symbol: 

 

 

Recommended separation distances between portable and mobile 

RF communications equipment and the EQUIPMENT or SYSTEM – 

For EQUIPMENT and SYSTEMS that are not LIFE – SUPPORTING 

Recommended  separation  distances  between  portable  and  mobile  RF  communications 
equipment and the [EQUIPMENT or SYSTEM] 

The [EQUIPMENT or SYSTEM] is intended for use in an electromagnetic environment in which 
radiated  RF  disturbances  are  controlled.  The  customer  or  the  user  of  the  [EQUIPMENT  or 
SYSTEM]  can  help  prevent  electromagnetic  interference  by  maintaining  a  minimum  distance 
between portable and mobile RF communications equipment (transmitters) and the [EQUIPMENT 
or  SYSTEM]  as  recommended  below,  according  to  the  maximum  output  power  of  the 
communications equipment. 

Rated maximum 

output power of 

transmitter 

Separation distance according to frequency of transmitter 

150 kHz to 80 MHz 

d = 1.16

p

 

80 MHz to 800 MHz 

d = 1.16

p

 

800 MHz to2.5 GHz 

d = 2.33

p

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according 
to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 

Содержание PHBPB16TL

Страница 1: ...ant Noted 6 4 2 Fitting the cuff 6 4 3 Body posture during measurement 7 4 4 Take Measurement 7 4 5 Data Transmission 8 4 6 Memory 9 5 Error Indication 9 6 Trouble Shooting 9 7 Specification 10 8 Abou...

Страница 2: ...o vibration such as ambulances and emergency helicopters A location where there is gas or flame A location where there is water or steam A location where chemicals are stored A location where the unit...

Страница 3: ...h several trials of testing to ensure the measurement accuracy The end user should conduct a manufacturer recommended inspection and calibration annually Blood pressure measurements determined with th...

Страница 4: ...Cuff Type BF Applied Part Model PHBPB16TL Applicable Arm Circumference 8 6 to 14 inches Note Symbol for TYPE BF APPLIED PART Display...

Страница 5: ...r demonstration purpose It is possible these batteries will therefore not last for 200 measurements The battery life can be confirmed in the bottom right of the display If the low battery symbol is di...

Страница 6: ...at can raise blood pressure Try not to take your blood pressure if you are nervous or upset If you are nervous anxious or agitated your blood pressure will rise Rest for 5 10 minutes before taking a r...

Страница 7: ...ensure the measurement accuracy 4 3 Body posture during measurement Relax place the elbow on the desk with palm facing up cuff should be at the heart level Fig 9 The reading accuracy may be affected...

Страница 8: ...ic pressure is greater or equal to 90mmHg as abnormal The display will flash to remind the end user of the blood pressure abnormality If the device caused any discomfort during measurement or fail to...

Страница 9: ...e memory mode and turn off 5 Error Indication List of Error code Error Cause How to correct Er 30 Power on cuff inflation rate is too low or main unit does not connect with the cuff 1 Reconnect the ai...

Страница 10: ...during the day which also will affect by the human emotional and physical condition 2 Record the variance and consult to the doctor If the above suggestion doesn t remediable please dial Pyle Technic...

Страница 11: ...Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible...

Страница 12: ...oltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable...

Страница 13: ...ions equipment and the EQUIPMENT or SYSTEM The EQUIPMENT or SYSTEM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of th...

Страница 14: ...axed person and still have HBP Let s look at the facts about blood pressure so you can better understand how your body works and why it is smart to start protecting yourself now no matter what your bl...

Страница 15: ...morning blood pressure and this happens in 50 of the cases Patients with morning hypertension have a 78 more chance of stroke compared with 48 of other hypertensive patients without morning high bloo...

Страница 16: ...ymbol for manufacturer 0123 Symbol for CE Symbol for Follow operating instructions Symbol for TYPE BF APPLIED PART Symbol for electrical and electronic equipment IP21 Symbol for the IP classification...

Отзывы: