on, no display. 

2

  If  still  cannot  power  on,  reinstall  the  batteries  or 

change new batteries 

Measured  value  are  abnormally 
high or low 

1

  Confirm the cuff is wrapped up correctly

 

2

  If  the  user  clothing  restricts  the  normal  flow,  please 

remove  the  obstructing  clothing  and  retake  the 
measurement 

3

  Relax, place the elbow on the desk with palm facing 

up;  cuff  should  align  with  heart  level.  Retake  the 
measurement 

Cuff  inflation  rate  is  too  low  or 
does not inflate 

1

 

Reconnect the air tube connector with the main unit 

2

 

Cuff  or  bladder  leakage,  if  necessary  purchased  a 
new one.   

Cuff deflates too quickly. 

1

 

Cuff  too  loose

confirm  the  cuff  is  wrapped  up 

correctly

 

Measure value is different from the 
hospital or the value is inconsistent 

1

 

Blood  pressure  value  is  varied  during  the  day  which 

also will affect by the human emotional and physical 
condition 

2

 

Record the variance and consult to the doctor 

*If the above suggestion doesn’t remediable, please dial Pyle Technical Support: 718-535-1800 

ext: 279 for help. 

 

7. Specification 

Description 

Arm automatic blood pressure monitor 

Model 

PHBPB16TL

 

 

Display 

LCD Digital Display 

Measuring 
principle 

Oscillometric Method 

Measurement 
Range 

Pressure: 0mmHg~280mmHg   

0kPa~37.3kPa

 

Pulse: 40 pulse/min ~180 pulse/min 

Accuracy 

Pressure : ±3mmHg         

±0.4kPa

 

Pulse: ±5%   

Memory 

90 sets memory of measurement values. 

Automatic 
power off 

Unattended 3 minutes 

Power source 

4 AAA Alkaline battery             

AC Adapter

AC 100~240 V

 

Battery Life 

Approx 200 measurements 

 Protection 
against  electric 
shock

 

   

Type BF

 

IP classification 

IP21 

Operating 
Environment 

Temperature

+41F

+104F

 

Humidity

≤93% 

Pressure: 70.0kPa

106.0kPa 

Altitude: 

  3 000 m 

Storage 

and 

transport 
Environment 

Temperature

-13F

+150F

 

Humidity

10%

95% 

Pressure:50.0kPa

106.0kPa 

Содержание PHBPB16TL

Страница 1: ...ant Noted 6 4 2 Fitting the cuff 6 4 3 Body posture during measurement 7 4 4 Take Measurement 7 4 5 Data Transmission 8 4 6 Memory 9 5 Error Indication 9 6 Trouble Shooting 9 7 Specification 10 8 Abou...

Страница 2: ...o vibration such as ambulances and emergency helicopters A location where there is gas or flame A location where there is water or steam A location where chemicals are stored A location where the unit...

Страница 3: ...h several trials of testing to ensure the measurement accuracy The end user should conduct a manufacturer recommended inspection and calibration annually Blood pressure measurements determined with th...

Страница 4: ...Cuff Type BF Applied Part Model PHBPB16TL Applicable Arm Circumference 8 6 to 14 inches Note Symbol for TYPE BF APPLIED PART Display...

Страница 5: ...r demonstration purpose It is possible these batteries will therefore not last for 200 measurements The battery life can be confirmed in the bottom right of the display If the low battery symbol is di...

Страница 6: ...at can raise blood pressure Try not to take your blood pressure if you are nervous or upset If you are nervous anxious or agitated your blood pressure will rise Rest for 5 10 minutes before taking a r...

Страница 7: ...ensure the measurement accuracy 4 3 Body posture during measurement Relax place the elbow on the desk with palm facing up cuff should be at the heart level Fig 9 The reading accuracy may be affected...

Страница 8: ...ic pressure is greater or equal to 90mmHg as abnormal The display will flash to remind the end user of the blood pressure abnormality If the device caused any discomfort during measurement or fail to...

Страница 9: ...e memory mode and turn off 5 Error Indication List of Error code Error Cause How to correct Er 30 Power on cuff inflation rate is too low or main unit does not connect with the cuff 1 Reconnect the ai...

Страница 10: ...during the day which also will affect by the human emotional and physical condition 2 Record the variance and consult to the doctor If the above suggestion doesn t remediable please dial Pyle Technic...

Страница 11: ...Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible...

Страница 12: ...oltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable...

Страница 13: ...ions equipment and the EQUIPMENT or SYSTEM The EQUIPMENT or SYSTEM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of th...

Страница 14: ...axed person and still have HBP Let s look at the facts about blood pressure so you can better understand how your body works and why it is smart to start protecting yourself now no matter what your bl...

Страница 15: ...morning blood pressure and this happens in 50 of the cases Patients with morning hypertension have a 78 more chance of stroke compared with 48 of other hypertensive patients without morning high bloo...

Страница 16: ...ymbol for manufacturer 0123 Symbol for CE Symbol for Follow operating instructions Symbol for TYPE BF APPLIED PART Symbol for electrical and electronic equipment IP21 Symbol for the IP classification...

Отзывы: