on, no display.
(
2
)
If still cannot power on, reinstall the batteries or
change new batteries
Measured value are abnormally
high or low
(
1
)
Confirm the cuff is wrapped up correctly
。
(
2
)
If the user clothing restricts the normal flow, please
remove the obstructing clothing and retake the
measurement
(
3
)
Relax, place the elbow on the desk with palm facing
up; cuff should align with heart level. Retake the
measurement
Cuff inflation rate is too low or
does not inflate
(
1
)
Reconnect the air tube connector with the main unit
(
2
)
Cuff or bladder leakage, if necessary purchased a
new one.
Cuff deflates too quickly.
(
1
)
Cuff too loose
,
confirm the cuff is wrapped up
correctly
。
Measure value is different from the
hospital or the value is inconsistent
(
1
)
Blood pressure value is varied during the day which
also will affect by the human emotional and physical
condition
(
2
)
Record the variance and consult to the doctor
*If the above suggestion doesn’t remediable, please dial Pyle Technical Support: 718-535-1800
ext: 279 for help.
7. Specification
Description
Arm automatic blood pressure monitor
Model
PHBPB16TL
Display
LCD Digital Display
Measuring
principle
Oscillometric Method
Measurement
Range
Pressure: 0mmHg~280mmHg
(
0kPa~37.3kPa
)
Pulse: 40 pulse/min ~180 pulse/min
Accuracy
Pressure : ±3mmHg
(
±0.4kPa
)
Pulse: ±5%
Memory
90 sets memory of measurement values.
Automatic
power off
Unattended 3 minutes
Power source
4 AAA Alkaline battery
AC Adapter
(
AC 100~240 V
)
Battery Life
Approx 200 measurements
Protection
against electric
shock
Type BF
IP classification
IP21
Operating
Environment
Temperature
:
+41F
~
+104F
;
Humidity
:
≤93%
Pressure: 70.0kPa
~
106.0kPa
Altitude:
≤
3 000 m
Storage
and
transport
Environment
Temperature
:
-13F
~
+150F
;
Humidity
:
10%
~
95%
Pressure:50.0kPa
~
106.0kPa