5.1.2 Sensor Cell Installation
1.
Remove the sensor cover (as required).
2.
Plug the sensor cell into the bottom of the transmitter.
3.
Replace the sensor cover.
NOTE:
If a new sensor cell is being installed, check for the presence of a
shorting plug and remove it before installation. See Section 3.2.1 for more
information.
Short Pin
Sensor Cell — Top View
5.2 Calibration
Model TX-KE and TX-KP gas detectors require periodic calibration with the
appropriate standard gas. PureAire Calibration Kits (optional) are recommended
for calibration purposes (see Section 5.4).
WARNING:
Before performing a calibration, it is the user’s responsibility to
confirm that they area is free of combustible gas.
IMPORTANT:
If the TX-KE or TX-KP is connected to an external alarm or
control system, the alarm or control system should be disabled or placed in a
standby mode during calibration to avoid an accidental alarm.
5.2.1 General
1.
Remove the cover of the calibration port (located on the left hand side of the
instrument housing).
2.
Remove the sensor cover.
3.
Place the calibration adapter over the sensor cell.