Product Information and User Instructions

Models CH5000S and CH5000

Caution – This device is only to be used by 
or on the prescription of a physician.   

Содержание CH5000

Страница 1: ...Product Information and User Instructions Models CH5000S and CH5000 Caution This device is only to be used by or on the prescription of a physician...

Страница 2: ...ning Introduction to Your Pure Oxygen Concentrator The Pure Oxygen Concentrator is designed as a stationary supplemental oxygen source A physician has prescribed supplemental oxygen because you are no...

Страница 3: ...s of the concentrator Rear View Figure B 7 Ventilation Slots 8 Humidifier bottle holder 9 Connection Adapter for humidifier bottle 10 Humidifier Bottle not included 11 Filter Door NORMAL OXYGEN ALARM...

Страница 4: ...of a physician or healthcare provider Oxygen supplied by this device should not be considered life supporting or life sustaining For those individuals who a healthcare professional has determined int...

Страница 5: ...the Oxygen Outlet Barb Figure D 5 Fill the humidifier bottle with distilled water 6 Using the strap firmly secure the humidifier bottle to the concentrator Figure E 7 Run the Connection Adapter under...

Страница 6: ...oximately every 12 months or as needed Please contact your equipment provider for a replacement Cannula Tubing and Humidifier Bottles Clean and replace as recommended and directed by the manufacturer...

Страница 7: ...unit has lost the power connection and the switch is in the on position See section A of the troubleshooting section of this manual Intermittent Beep No power to Unit See section A of the trouble sho...

Страница 8: ...in back are blocked Check the rear cooling vents are not blocked or restricted If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned this was the issue 3 Blo...

Страница 9: ...ws ranging 0 5 to 5 LPM Outlet Pressure 5 5 PSI 0 5 PSI Flow Range 0 5 to 5 LPM Low Flow Alarm 0 5 LPM Pressure Relief 35 PSI 5 PSI 247kPa 35 kPa Change in Maximum Recommended Flow when Back Pressure...

Страница 10: ...rtable and mobile RF communications transmitters and the Pure Oxygen Concentrator as recommended below According to the maximum output power of the communication equipment Separation distance accordin...

Страница 11: ...than a Drive Medical Authorized Service Professional shall immediately void this warranty Drive Medical will not accept damages or charges for labor parts or expenses incurred in making field repairs...

Страница 12: ...Drive Medical 99 Seaview Blvd Port Washington NY 11050 Toll free 877 224 0946 www drivemedical com...