DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000
.......................................................................................................................
8
Symptom
Probable Cause
Potential Action
A. Unit does not operate.
No lights are seen on
the front panel when the
power switch is “On”. An
audible alarm is heard
(intermittent beep).
1. Power cord
not properly
inserted into
wall outlet.
Check power connection at the wall outlet.
2. Issue with
power supplied
to wall outlet.
Check your home circuit breaker and reset if
necessary. Try using a different wall outlet if
the situation occurs again.
B. Unit is running and
either the red and green
light OR red and yellow
light is lit when the pow-
er switch is in the “On”
position. An audible
alarm is heard.
1. Air filter is
blocked.
Check the air filters. If dirty, clean/replace
the filter(s) per the instructions in this manual
(see "Cleaning and Maintenance" section
of this manual). If unit ceases to alarm after
a minimum of two minutes and proper flow
appears to have returned, this was the issue.
2. Vents in back
are blocked.
Check the rear cooling vents are not blocked
or restricted. If unit ceases to alarm after a
minimum of two minutes and proper flow
appears to have returned, this was the issue.
3. Blocked
tubing
Detach cannula, catheter, or face mask from
oxygen outlet. If unit ceases to alarm after
a minimum of two minutes and proper flow
appears to have returned, this was the issue.
Check for obstructions or kinks in external
cannula, catheter, face mask, or tubing.
Clean or replace, if necessary.
4. Blocked /
defective humid-
ifier bottle.
Detach the humidifier from the oxygen outlet.
If unit ceases to alarm after a minimum of
two minutes and proper flow appears to
have returned, this was the issue.
5. Flow meter
set too low.
Turn flow meter dial counter-clockwise to a
minimum of 0.5 LPM or greater. If unit ceas-
es to alarm after a minimum of two minutes
and proper flow appears to have returned,
this was the issue.
C. If any other problems
occur with your oxygen
concentrator or none of
the solutions above work
Turn your unit “OFF”, switch to your reserve
/ alternate oxygen system, and contact your
service provider immediately.
Basic Troubleshooting