DRIVE PURE OXYGEN CONCENTRATOR  MODELS CH5000S and CH5000

.......................................................................................................................

8

Symptom

Probable Cause

Potential Action

A. Unit does not operate. 
No lights are seen on 
the front panel when the 
power switch is “On”. An 
audible alarm is heard 
(intermittent beep).

1. Power cord 
not properly 
inserted into 
wall outlet.

Check power connection at the wall outlet. 

 

2. Issue with 
power supplied 
to wall outlet.

Check your home circuit breaker and reset if 
necessary. Try using a different wall outlet if 
the situation occurs again.

B. Unit is running and 
either the red and green 
light OR red and yellow 
light is lit when the pow-
er switch is in the “On” 
position.  An audible 
alarm is heard.

1. Air filter is 
blocked.

Check the air filters. If dirty, clean/replace 
the filter(s) per the instructions in this manual 
(see "Cleaning and Maintenance" section 
of this manual).  If unit ceases to alarm after 
a minimum of two minutes and proper flow 
appears to have returned, this was the issue.

 

2. Vents in back 
are blocked.

Check the rear cooling vents are not blocked 
or restricted.  If unit ceases to alarm after a 
minimum of two minutes and proper flow 
appears to have returned, this was the issue.

 

3. Blocked 
tubing

Detach cannula, catheter, or face mask from 
oxygen outlet.  If unit ceases to alarm after 
a minimum of two minutes and proper flow 
appears to have returned, this was the issue.  
Check for obstructions or kinks in external 
cannula, catheter, face mask, or tubing.  
Clean or replace, if necessary.

 

4. Blocked / 
defective humid-
ifier bottle.

Detach the humidifier from the oxygen outlet. 
If unit ceases to alarm after a minimum of 
two minutes and proper flow appears to 
have returned, this was the issue.

5. Flow meter 
set too low.

Turn flow meter dial counter-clockwise to a 
minimum of 0.5 LPM or greater.  If unit ceas-
es to alarm after a minimum of two minutes 
and proper flow appears to have returned, 
this was the issue.

 C. If any other problems 
occur with your oxygen 
concentrator or none of 
the solutions above work

Turn your unit “OFF”, switch to your reserve 
/ alternate oxygen system, and contact your 
service provider immediately.

Basic Troubleshooting 

Содержание CH5000

Страница 1: ...Product Information and User Instructions Models CH5000S and CH5000 Caution This device is only to be used by or on the prescription of a physician...

Страница 2: ...ning Introduction to Your Pure Oxygen Concentrator The Pure Oxygen Concentrator is designed as a stationary supplemental oxygen source A physician has prescribed supplemental oxygen because you are no...

Страница 3: ...s of the concentrator Rear View Figure B 7 Ventilation Slots 8 Humidifier bottle holder 9 Connection Adapter for humidifier bottle 10 Humidifier Bottle not included 11 Filter Door NORMAL OXYGEN ALARM...

Страница 4: ...of a physician or healthcare provider Oxygen supplied by this device should not be considered life supporting or life sustaining For those individuals who a healthcare professional has determined int...

Страница 5: ...the Oxygen Outlet Barb Figure D 5 Fill the humidifier bottle with distilled water 6 Using the strap firmly secure the humidifier bottle to the concentrator Figure E 7 Run the Connection Adapter under...

Страница 6: ...oximately every 12 months or as needed Please contact your equipment provider for a replacement Cannula Tubing and Humidifier Bottles Clean and replace as recommended and directed by the manufacturer...

Страница 7: ...unit has lost the power connection and the switch is in the on position See section A of the troubleshooting section of this manual Intermittent Beep No power to Unit See section A of the trouble sho...

Страница 8: ...in back are blocked Check the rear cooling vents are not blocked or restricted If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned this was the issue 3 Blo...

Страница 9: ...ws ranging 0 5 to 5 LPM Outlet Pressure 5 5 PSI 0 5 PSI Flow Range 0 5 to 5 LPM Low Flow Alarm 0 5 LPM Pressure Relief 35 PSI 5 PSI 247kPa 35 kPa Change in Maximum Recommended Flow when Back Pressure...

Страница 10: ...rtable and mobile RF communications transmitters and the Pure Oxygen Concentrator as recommended below According to the maximum output power of the communication equipment Separation distance accordin...

Страница 11: ...than a Drive Medical Authorized Service Professional shall immediately void this warranty Drive Medical will not accept damages or charges for labor parts or expenses incurred in making field repairs...

Страница 12: ...Drive Medical 99 Seaview Blvd Port Washington NY 11050 Toll free 877 224 0946 www drivemedical com...

Отзывы: