DRIVE PURE OXYGEN CONCENTRATOR  MODELS CH5000S and CH5000

.......................................................................................................................

6

Cleaning and Maintenance 

Unplug the concentrator before cleaning.  Do not use oil or grease.  

Concentrator
Periodically, with a damp cloth, wipe the outside of the concentrator using a solution of hot 
water and dishwashing detergent.  If any medical cleaning solutions are used, be sure to 
follow the manufacturer’s directions.

Filters
The Foam Intake filter is a washable filter.  Rinse and wash as needed.
1)  Remove the filter cover.
2)  Remove both the Foam Intake filter and the HEPA filter.
3)  Separate the Foam Intake filter from the HEPA filter.
4)  Wash in a solution of warm water and dishwashing detergent.
5)  Thoroughly rinse and dry.
6)  Replace Foam Intake filter, HEPA filter, and Filter Cover.

The HEPA Intake Filter is a replaceable filter.  Replacement is 
recommended approximately every 12 months, or as needed.  
Please contact your equipment provider for a replacement.       

Cannula, Tubing, and Humidifier Bottles 
Clean and replace as recommended and directed by the 
manufacturer or your equipment provider.  

Model, Parts and Accessories

Model Description
CH5000S 

Pure Oxygen Concentrator, with Oxygen Sensor

CH5000 

Pure Oxygen Concentrator, without Oxygen Sensor

Parts Description
PL-2001844 

Hooded Swivel Caster 

1/each

PL-2001853 

HEPA Intake Filter 

1/each

PL-2001854 

Reusable Foam Intake Filter 

1/each

PL-3000706 

Filter Intake Door 

1/each

The following humidifier bottle is recommended for use with the Pure Oxygen Concentrator.  

Accessories Description
HUM 001 

Humidifier Bottle 

50/case

HUM-TUB 

Humidifier Adapter Tubing 

50/case

CON 550 

Ridged Oxygen Tubing Connector 

50/case

SOFT 225 

Non-kinking Cozy Nasal Cannula 

25' 25/case 

SOFT 250 

Non-kinking Cozy Nasal Cannula 

50' 20/case

TUB NK 25 

Non-kinking Oxygen Tubing 

25'  25/case

TUB NK 50 

Non-kinking Oxygen Tubing 

50' 20/case

Filter Door

HEPA

Foam Intake

Содержание CH5000

Страница 1: ...Product Information and User Instructions Models CH5000S and CH5000 Caution This device is only to be used by or on the prescription of a physician...

Страница 2: ...ning Introduction to Your Pure Oxygen Concentrator The Pure Oxygen Concentrator is designed as a stationary supplemental oxygen source A physician has prescribed supplemental oxygen because you are no...

Страница 3: ...s of the concentrator Rear View Figure B 7 Ventilation Slots 8 Humidifier bottle holder 9 Connection Adapter for humidifier bottle 10 Humidifier Bottle not included 11 Filter Door NORMAL OXYGEN ALARM...

Страница 4: ...of a physician or healthcare provider Oxygen supplied by this device should not be considered life supporting or life sustaining For those individuals who a healthcare professional has determined int...

Страница 5: ...the Oxygen Outlet Barb Figure D 5 Fill the humidifier bottle with distilled water 6 Using the strap firmly secure the humidifier bottle to the concentrator Figure E 7 Run the Connection Adapter under...

Страница 6: ...oximately every 12 months or as needed Please contact your equipment provider for a replacement Cannula Tubing and Humidifier Bottles Clean and replace as recommended and directed by the manufacturer...

Страница 7: ...unit has lost the power connection and the switch is in the on position See section A of the troubleshooting section of this manual Intermittent Beep No power to Unit See section A of the trouble sho...

Страница 8: ...in back are blocked Check the rear cooling vents are not blocked or restricted If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned this was the issue 3 Blo...

Страница 9: ...ws ranging 0 5 to 5 LPM Outlet Pressure 5 5 PSI 0 5 PSI Flow Range 0 5 to 5 LPM Low Flow Alarm 0 5 LPM Pressure Relief 35 PSI 5 PSI 247kPa 35 kPa Change in Maximum Recommended Flow when Back Pressure...

Страница 10: ...rtable and mobile RF communications transmitters and the Pure Oxygen Concentrator as recommended below According to the maximum output power of the communication equipment Separation distance accordin...

Страница 11: ...than a Drive Medical Authorized Service Professional shall immediately void this warranty Drive Medical will not accept damages or charges for labor parts or expenses incurred in making field repairs...

Страница 12: ...Drive Medical 99 Seaview Blvd Port Washington NY 11050 Toll free 877 224 0946 www drivemedical com...

Отзывы: