FLOWPULSE INSTRUCTION MANUAL
36
Setup Menu
This menu enables setting up of the RS232 port. The default setting when
interfacing using FlowPulse PC is 57600 Baud 8bits-No Parity-1Stop bit.
There is normally no need to change this as FlowPulse PC should
automatically detect a FlowPulse connected however a specific comport can
be selected in the ‘Serial Port’ sub menu.
The ‘Recording Interval’ menu controls the time interval before a new flow
record and trace is saved to file when the save-to-feature is in use.
The ‘Clear Flow Record Chart’ option allows the Flow Record chart to be
cleared.
The ‘Work Folder’ allows the setting of a default folder for data files to be
saved.
System Menu
This menu provides the option to perform the following actions:
OPTION
DECRIPTIONS
Save Device Parameters
Extract configuration parameters from the device
and save onto PC as a parameter file. This is saved
to FlowPulse PC’s installation folder and the
filename is prefixed with “ParamFlowPulse
-
”.
Load Parameter to Device
Load the parameters from a parameter file onto
FlowPulse. This should only be used with a file
that has a prefix of “ParamFlowPulse
-
”
Bootloader Control
Launch the Bootloader PC program for connecting
to FlowPulse
’s Bootloader. This allows the
firmware on FlowPulse to be upgraded. Note that
any customised parameters will be erased during a
firmware upgrade, and therefore it is
recommended that device parameters are saved
onto PC before performing a firmware upgrade.
Reset Device to Bootloader
Reset FlowPulse into Bootloader mode. The
Bootloader PC program must be used to connect
to FlowPulse in this mode. FlowPulse will resume
normal operation if no connection is established
within 30 seconds.
Parameter Factory Reset
This results all parameters to factory default.