![Pulsar Measurement IMP Lite Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/pulsar-measurement/imp-lite/imp-lite_instruction-manual_1625416008.webp)
IMP LITE INSTRUCTION MANUAL
8
Product Specification
PHYSICAL
Dimensions
O/A height186mm
x O/A diameter 133mm (5.24” x
7.32”)
Mounting
1.5” BSP/NPT (IMP3 and 6 models) 2” BSP/NPT
(IMP10)
Weight
Nominal 1 kg
Cable entry detail
2 off cable glands 4.5
–
10mm. (torque to 2NM)
ENVIRONMENTAL
IP Rating (Wall)
IP67
Max. & min. temperature
(Ambient)
-40ºC to +65ºC (-4ºF to 149ºF)
Max. & min. temperature
(Transducer)
-40ºC to +85ºC (-4ºF to 176ºF)
CE Approval
See EU Declaration of Conformity
PERFORMANCE
Beam angle (-3dB half power)
<10° inclusive
Operating frequency
Lite 3 125kHz, Lite 6 75kHz, Lite 10 41kHz
Range
Lite 3 0.2
–
3m (0.66
–
10ft)
Lite 6 0.3
–
6m (0.98
–
20ft)
Lite 10 0.3
–
10m (0.98
–
33ft)
Current output
3.5
–
22mA
Accuracy
± 0.25% or 6mm (0.24 inches) whichever is greater
Resolution
± 0.1% or 2mm (0.08inches) whichever is greater
4-20mA output
Resolution 5
µA
Temperature compensation
Via internal temperature sensor (± 0.5°C accuracy)
Response time
Power up to stable reading = 10 seconds
PROGRAMMING
PC Programming
Via RJ11 port to RS232 using IMP PC software
On-board programming
Via 4 button keypad and a 4-digit LCD display
SUPPLY
Power supply
11 - 30V DC
Pulsar Measurement Limited operates a policy of constant development and
improvement and reserve the right to amend technical details, as necessary.