Electrical data
NOTE
The X-ray system is in need of the following power supply (see table „Power supply
Generator).
Type generator
PROVARIO
HF 50
PROVARIO
HF 60
PROVARIO
HF 80
Output Power
50kW
65kW
80kW
Power supply voltage
400V AC
Phase
3PH-N-PE
Power frequency
50/60 Hz
Electrical resistance per
phase
0,3Ω
0,2Ω
0,12Ω
Fuse
50A
List (Power supply generator)
Type generator
Venus 50R, 3
phase
Venus 50R, 1
phase
Venus 32R, 3
phase
Venus 32R, 1
phase
Output Power
50kW
50kW
32kW
32kW
Power supply
voltage
380V AC/400V
AC
220V AC
380V AC /
400V AC
220V AC/230V
AC
Phase
3PH-N-PE
1phase
3PH-N-PE
1phase
Power frequency
50/60Hz
50/60Hz
50/60 Hz
50/60 Hz
Electrical
resistance per
phase
0,15 Ω/0,17 Ω
0,5Ω
0,27 Ω/0,29 Ω
0,5Ω
Fuse
63A
16A
63A
16A
List (Power supply generator)
Type generator
CMP 200 DR
50
CMP 200 DR
60
CMP 200 DR
80
Output Power
50kW
65kW
80kW
Power supply
voltage
400/480 V AC only
Phase
3PH-N-PE
Power frequency
50/60 Hz
Electrical
resistance per
phase
0,17Ω/0,24Ω
0,13Ω/0,19Ω
0,10 Ω/0,15Ω
Fuse
60A/480V
List (Power supply generator)
Type generator
RFX 50
RFX 60
RFX 80
Output Power
55kW
65kW
80kW
Power supply
voltage
380V/400V/440V/480V
Phase
3PH-PE (N not needed)
Power frequency
50/60 Hz
Electrical
resistance per
phase
0,15Ω
0,17Ω
0,20Ω
0,24Ω
0,15Ω
0,17Ω
0,20Ω
0,24Ω
0,10Ω
0,11Ω
0,14Ω
0,16Ω
Fuse
50A/480V