PRS 500 B User Manual
5014-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
19 from 42
4.1.1.6
Exposure preparation / exposure release
•
At the X-ray generator operator console control panel, select the desire X-ray equipment (bucky
table with bucky unit) by selecting the corresponding X-ray equipment/exposure technique
switch.
•
Press the desired Anatomically Programmed Radiography (APR) switch or manually select the
desired exposure factors. Start the exposure by pressing the switches for exposure
preparation/exposure release.
4.1.1.7
On table exposures
•
Place a cassette to the desired position on the table top.
•
Move X-ray tube to the desired position and adjust SID.
•
Press the collimator light switch (button 9, figure collimator) to turn on the collimator light, and
view the opening of the collimator shutters in both axes relatives to the cassette and object size.
•
Place object on cassette.
•
Adjust the light field with the adjuster 1 and 7 (figure Collimator) onto the size of the used cassette.
So the radiation field will be limited to the size of the cassette.
•
At the X-ray generator operator console control panel, select the desired X-ray equipment (bucky
table without bucky unit) by selecting the corresponding X-ray equipment/exposure technique
switch.
•
Press the desired Anatomically Programmed Radiography (APR) witch or manually select the
desired exposure factors. Start the exposure by pressing the switches for exposure preparation and
exposure release.
4.1.1.8
Exposures with the lateral detector holder (optional)
•
1. Step:
Move the tube down beside the table