Function of the PRS 500 B
4.6.1
Switching the PRS 500 B ON
(1)
Switch on the main switch of the X-ray room
(2)
Switch on the operation PC (The PC starts up normally)
(3)
Switch on the EC-Box at the air switch on the side of the EC-Box. The green light on top of the EC-
Box lights up and shows the on-state.
(4)
Press the “ON”-button (appr. 2 seconds) on the control console. The LED ring around the button
lights up and shows the on-state of the system. The Touch-PC on the tube head starts up and the
software will also start automatically.
(5)
After the two PCs have finished the start-up process the system is ready to use.
4.6.2
Switching the PRS 500 B OFF
CAUTION!
Before switching off the PRS 500 B system ensure that the tube head is turned
to 0° to avoid accidently turning movements and damages of the tube head.
(1)
Press the “OFF”-button (appr. 2 seconds) on the control console.
The LED ring around the button lights up and shows the turning off state of the system. After a few
seconds the power of the electronic is turned off and the system is ready to be switched off.
NOTE
To perform a regular shutdown of the Touch PC it is necessary that the CONAXX
software is running on the operator PC and on the Touch PC at the same time while
the “OFF” button on the control console is pressed. If one component is not working
there is no communication between both parts and the system cannot shut down.
(2)
Switch off the air switch on the side of the EC-Box.
(3)
Exit the software of the operation PC.
(4)
Switch off the main switch of the X-ray room.
NOTE
If CONAXX is closed on the Touch PC or operator PC before the system is shut down
it is possible to perform a “hard shutdown”.
The “OFF” button on the control console has to be pressed 5 seconds at least to cut
the power supply to the system.