PRS 500 B User Manual
5014-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
23 from 42
4.1.4
Operation X-ray Tube
NOTE
The X-Ray tube needs to be warmed-up daily in order to extend the life of the tube
and prevent tube arcs (Especially when the X-Ray tube was not used for a long
period). The seasoning procedure shall be done upon turning on the generator for
the first time.
Follow X-Ray tube manufacturer´s recommended seasoning procedure.
If X-Ray tube manufacturer´s seasoning is not available, then use the following
procedure:
Set Generator: Large focal spot, 200mA, 40mAs
Take 8 exposures starting at 50 kV and increment the kV steps of 10 kV up to 120 kV
(Exposure every 30 seconds, otherwise tube may arc).
See User Manual of the generator and CONAXX 2 User Manual 5.3.
Detailed information please find in the enclosed User Manual of the X-ray tube.
4.2
Operation X-ray generator
Detailed information please find in the enclosed User Manual of the generator.
4.3
Operation Bucky, Grid entity
Detailed information please find in the enclosed User Manual.
4.4
Operation RAPIXX system
Detailed information please find in the enclosed User Manual.
4.5
Operation Software
Detailed information please find in the enclosed installation- and User Manual CONAXX 2.