Protean Instrument Corporation
79
Data MPC
900
The top left button allows you to toggle between viewing the beta cpm and efficiency.
The + and – buttons allow fine control of the cursor (vertical green bar) to look at various
points. Pressing the vertical blue bars causes the cursor to snap to that line if there are
points at that location, otherwise the button presses are ignored. This feature
eliminates the need for mu and – button presses to get to the beginning or end of
the data set. Below the + and – buttons are data for the current cursor position.
Pressing the
DELETE
button deletes the point at the cursor location.
Pressing the
SETTINGS
button allows viewing of the control chart settings.
The first line is the control chart name. The second line is the counting mode used
during acquisition. The third line is the amount of time for each acquisition.
The date range is the date of the first limit point to the date of the last limit point that is
used in the Beta Mean and Standard Deviation values which are shown on the next four
lines. The last three lines show the check source information.