Turn on the PDS-510 by sliding the Main Power switch to the ON position on the side of the
meter.
6.
Switch the POWER to the ON position.
7.
Switch the RANGE to the HIGH position.
8.
Switch the HIGH VOLTAGE to the ON position.
9.
By using the knob, adjust the voltage to 1.00 (1,000 volts).
10.
In the PicoScope software, if needed, move the yellow trigger icon above the scope’s zero line
(0.2A) on the left vertical scale (Figure 7).
11.
Ensure that the scope horizontal scale is set to 100 ns per division (Figure 7).
12.
Without inserting a shielded bag onto the fixture, set the Discharge Arm onto the Capacitive
Arm. Set the 5 lbs electrode on top of the Discharge Arm.
Figure 7. Screenshot of the PicoScope software with the “Waveform Verification Setup file” open.