8
4.
Connect the current probe coming from the Capacitive Arm to the male end of the 50-ohm
terminator (Figure 5).
5.
Connect the female end of the 50-ohm terminator to Channel A of the PicoScope (Figure 6).
6.
Connect the supplied USB cable from the PicoScope unit to a computer.
Figure 5. The 50 ohm terminator connects to the Tektronix®
Figure 6. Connect the female end of the 50-ohm
CT1 Current Probe
terminator to Channel A of the PicoScope
04 SYSTEM VERIFICATION TEST
Software Setup
1.
Install the PicoScope software onto the computer by downloading it from Pico’s website at
https://www.picotech.com/downloads. Use the version that is closest to the model of your
Pico Scope.
2.
Insert the supplied USB Thumb Drive and copy the 3 files named “Waveform Verification
Setup“, “Shielding Bag Test File” and “Shielding Bag Test File with Manual Cursors” to your
computer’s desktop.
3.
Launch the PicoScope software. Note: It is recommended that you launch the Pico software
first and then open the files from the thumb drive.
4.
From the PicoScope software, click on File > Open and select the “Waveform Verification
Setup file” from the desktop