Error indications
Readout at hardware error
Error search
Readout
Cause
Test of internal CJC sensor
CJ.ER
CJC sensor defect or temperature
outside range
Checksum test of the configuration
in FLASH
FL.CO
Error in FLASH
Check measurement of analog
output current
AO.ER
1) No load on the current output
(only S4…20 / S20…4 mA)
Communications test
UMU-FD / UMU 100
NO.CO Connection error
Check that Input signal matches
Input configuration
IN.ER
1) Error levels on Input
Check that saved configuration in
UMU 100 matches device
TY.ER
Configuration is not UMU 100
All error indications in the display flash once per second. The help text
!
explains the error. If the error is a sensor error, the display flahes as well
–
this is acknowledged (stopped) by pushing the
button.
1)
The error is reset by switching off and then switching on the supply voltage
to the device.