![promesstec UMU 100-A-B Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/promesstec/umu-100-a-b/umu-100-a-b_user-manual_1616639016.webp)
16
promesstec GmbH
I Niedersachsenstraße 4 I D- 48465 Schüttorf I Tel.:+49 (0)5923/ 90 229 0 I Fax:+49 (0)5923/ 90 229 29
E-Mail: [email protected] I Internet: www.promesstec.com
Order
UMU 100 = Universal-measuring transducer
UMU-FD = Display- / programming front
UMU-CJC = CJC-connector
10. Electrical specifications
Environmental conditions
Specification area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -20°C to +60°C
Storage temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . -20°C to +85°C
Calibration temperature . . . . . . . . . . . . . . . . . . . . . . . . . .20...28°C
Relative humidity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < 95% RF (non cond.)
Protection degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP20
Installation in pollution degree 2 & measurement / overvoltage category II
Mechnical specification
Dimensions (HxWxD) . . . . . . . . . . . . . . . . . . . . . . . . . . 109 x 23,5 x 104 mm
Dimensions (HxWxD) with UMU-FD. . . . . . . . . . . . . . . .109 x 23,5 x 116 / 131 mm
Weight, approx. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155 g
Weight with UMU-FD (approx.). . . . . . . . . . . . . . . . . . . .170 g / 185 g
DIN rail type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DIN EN 60715/35 mm
Cable cross section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0,13...2,08 mm2 / AWG
26...14 stranded wire
Screw terminal torque. . . . . . . . . . . . . . . . . . . . . . . . . . . 0,5 Nm
Vibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 60068-2-6
2...13,2 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 mm
13,2...100 Hz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±0,7 g
Common electrical specifications
Supply voltage, universal . . . . . . . . . . . . . . . . . . . . . . . . 21,6...253 VAC, 50...60 Hz
or 19,2...300 VDC
Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 mA SB / 250 VAC
Max. required power.
. . . . . . . . . . . . . . . . . . . . . . . . . . . ≤ 2,0 W
Max. power dissipation. . . . . . . . . . . . . . . . . . . . . . . . . .
≤ 2,0 W
Isolation voltage, test. . . . . . . . . . . . . . . . . . . . . . . . . . . .2,3 kVAC
Isolation voltage, working . . . . . . . . . . . . . . . . . . . . . . . .250 VAC (reinforced) / 500
VAC (basic)
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . UMU-FD
Signal dynamics, input / output . . . . . . . . . . . . . . . . . . . .24 Bit / 16 Bit
Signal- / noise ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 60 dB (0...100 kHz)
Responsing time (0...90%, 100...10%):
Temperature Input.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≤ 1 s
mA- / V-input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
≤ 400 ms
Auxiliary supplies
2-wire-supply (terminal 44...43) . . . . . . . . . . . . . . . . . . . .25...16 VDC / 0...20 mA