11. Visualisaton in the UMU-FD
Sensor error detection and input signal outside range
Sensor error check:
Device
Configuration
Sensor error detection:
UMU 100-A-B
OUT.ERR=NONE.
OFF
Else:
ON
Outside range readout (IN.LO, IN.HI):
If the valid of the A/D converter or the polynomial is exceeded
Input
Range
Readout
Limit
Volt
0…1 V / 0,2…1 V
IN.LO
< - 25 mV
IN.HI
> 1,2 V
0…10 V / 2…10 V
IN.LO
< - 25 mV
IN.HI
>12 V
CURR
0…20 mA / 4…20 mA
IN.LO
< - 1,05 mA
IN.HI
> 25.05 mA
LIN.R
0…800Ω
IN.LO
< 0 Ω
IN.HI
> 1075 Ω
0…10 kΩ
IN.LO
< 0 Ω
IN.HI
< 110 kΩ
POTM
-
IN.LO
< - 0,5 %
IN.HI
> 100,5 %
TEMP
TC / RTD
IN.LO
< temperature range -2°C
IN.HI
> temperature range +2°C
Display readout below min. / above max. (-1999, 9999):
Input
Range
Readout
Limit
All
All
-1999
Display readout <-1999
9999
Display readout > 9999
Sensor error detection limits
Sensor error detection (SE.BR, SE.SH):
Input
Range
Readout
Limit
CURR
Loop break (4…20 mA)
SE.BR
<= 3.6 mA; > = 21 mA
POTM
All, SE.BR on all 3-wire
SE.BR
> ca. 126 k
Ω
LIN.R
0…800 Ω
SE.BR
> ca. 875
Ω
0…10 kΩ
SE.BR
> ca. 11 k
Ω
TEMP
TC
SE.BR
> ca. 750 k
Ω / (1.25 V)
RTD, 2-, 3-, and 4-wire
No SE.SH for Cuxx, Pt10,
PT20 and Pt50
SE.BR
> Ca. 15 k
Ω
SE.SH
< ca. 15 k
Ω
UMU 100