VERY IMPORTANT
In order to measure digital channels correctly, previously, the channel must be
defined as digital (see Edit Channel Plan function at section '4.2.1 Global
Configuration menu) and, if necessary, the CHANNEL BW parameter should be
redefined by means of the Channel-Frequency configuration menu. When a
channel has been defined as digital, the PROMAX-10
Premium
tunes it at its
central frequency.
Digital channels power measurement is carried out by means of an integration
method. The
PROMAX-10
Premium
divides the bandwidth of the channel (
CHANNEL
BW
) in sections of 230 kHz (4 by each MHz approximately) and measures the
contribution of each one to the total power of the channel. This way, a very exact
measurement is obtained, specially in the case of degraded channels, because the
channel flatness is had in consideration.
Figure 20.-
Measuring the power of a digital channel by integration.
The
C/N ratio
measurement shows the ratio between the digital channel power
and the noise power. The user can define the frequency where the noise power must be
measured. Two possibilities exist:
I. Absolute method
Selecting in the Channel-Frequency configuration menu the
NOISE MODE
parameter as
FREQ
, the equipment will interpret the
NOISE FREQ.
value as the
frequency where the noise power measurement must be done. Naturally the user must
be sure that the
NOISE FREQ.
value corresponds with a free channel.
II. Relative method
Selecting in the Channel-Frequency configuration menu the
NOISE MODE
parameter as
F
, the unit will make the measurement of noise at the frequency obtained
to add to the tuning frequency (channel central frequency) the value defined for the
NOISE FREQ.
parameter. The
PROMAX-10
Premium
takes by default
NOISE FREQ.
= BW/2 + 0.5 MHz
, where
BW
is the channel bandwidth defined in the channel plan,
thus for example, if it is desired to measure the C/N ratio of a digital channel with a
bandwidth (BW) of 8 MHz,
NOISE FREQ.
will take as value 4.5 MHz. The Channel-
Frequency configuration menu allows the user to redefine this parameter; when tuning a
new channel the equipment will return to assign to
NOISE FREQ.
the default value.
Содержание 10 SE
Страница 1: ...PROMAX 10 Premium ANALIZADOR DE TV CABLE CABLE TV ANALYSER ANALYSEUR DE TV C BLE 0 MI1617...
Страница 2: ......
Страница 4: ......
Страница 5: ...SUMARIO CONTENTS SOMMAIRE Manual espa ol English English manual Fran ais Manuel fran ais...
Страница 6: ......
Страница 16: ...MANUAL DE INSTRUCCIONES PROMAX 10 Premium P gina 8 11 2008...
Страница 56: ...MANUAL DE INSTRUCCIONES PROMAX 10 Premium P gina 48 11 2008...
Страница 58: ...MANUAL DE INSTRUCCIONES PROMAX 10 Premium P gina 50 11 2008...
Страница 106: ...USER S MANUAL PROMAX 10 Premium Page 46 11 2008...
Страница 154: ......