Page 18
Press the symbols
and
to open the
calibration function. Trigger the calibration by
pressing
.
The “Calibration Result”
window displays the measured values as well
as the deviation from the expected values in
percent. Successful calibration is indicated
through “Calibration OK”. Confirm the
calibration by pressing
and to
return to the measuring mode.
Calibration results with deviations greater
than 15 % are indicated with the message
“Calibration is not OK”. This can be caused
by a soiled or damaged standard or a soiled
optic window. It is suggested to reject such a
calibration result by pressing
before
determining the reason for the deviation.
Clean the standard as described in chapter
15.2
“Cleaning” on page 44 and recalibrate.
When repeatedly experiencing a deviation greater than 15%, service or factory
calibration might be required. Please contact Proceq or your authorized
Proceq agent for further information.
The standard settings for R can be seen by
pressing
,
and
.
Remark:
You can change the calibration interval as
described in chapter 13.6.8
Содержание ZRP 6030+
Страница 1: ...ZRP 6030 Retroreflectometer Instruction Manual...
Страница 35: ...Page 35 13 Menu 13 1 Menu structure...
Страница 51: ......