Page 40
Database
Path:
->
and selecting “Setup”
Select
“Database”. Here it is possible to
create a new database or select, rename or
delete an existing one. You can export a
database to a *.csv (comma separated text)
file, which can be edited e.g. in Excel.
Before deleting a database, make sure
you have selected the correct one and
made a backup of it if necessary.
Printer
Path:
->
and selecting “Setup”
Select
“Printer”. Select an available printer
from the USB or Thermal printer list to
determine which information should be
included on reports with the selected printer.
On both lists you can select the following
information:
Location, Street, Position, Section, Direction,
Line, Type, Remarks, Producer, Product,
Manufacture-no, Certificate, Glass beads,
Type, Mixture, Batch no, Supplier, Product,
Type, Batch no
Calibrate
Path:
->
and selecting “Setup”
Select
“Calibration” to change general
calibration settings.
“Calibration interval” sets the time after which
the device will require calibration on the
standard.
“Show deviation 2nd std” determines whether
the deviation will be shown when calibrating
with the second standard.