Page 26
Always charge the battery with the charger supplied.
If possible, the battery should always be charged completely which is
shown with the green charging status indication light.
The battery may only be replaced by Proceq or by an authorized Proceq-
agent.
If you have not used your ZRM 6006 during several months, the battery
should be charged before using.
9
Maintenance
9.1 Maintenance carried out by the user
You may only carry out the following maintenance and repair yourself:
•
Charging as described in chapter 8.3 on page 25.
•
Cleaning as described in chapter 9.2 on page 26.
All other maintenance and repair work may only be carried out by Proceq SA
or your authorized Proceq agent, otherwise all warranty voids.
Make sure that the ZRM 6006 is unplugged from power supply and turned off
before any maintenance, except for charging.
The instrument consists of sensitive optic and electronic precision parts.
Protect it from shocks, moisture and dust. Please store the instrument in the
transportation case.
9.2 Cleaning
We recommend that the instrument is checked and certified by Zehntner every two
years. Otherwise clean the instrument periodically using water- and oil-free
compressed air (max. 1.5 to 2 bar), or wipe the exit window on the bottom of the
optics housing with a soft brush.
The exit window is coated with a special antireflection coating that should not be
damaged. If the soft brush is not sufficient for cleaning the window, you may use a
soft paper tissue and window cleaning liquid.
The optics housing may not be opened under any circumstances since the
measuring geometry would be misaligned. The measuring geometry can only
be adjusted by means of special testing equipment in our factory.