Page 25
8
Built-in battery and charging
8.1 Battery
The used Li-Ion-Mn battery has a very high capacity. The capacity is continuously
being observed. When the voltage becomes too weak the battery sign appears
on the display and the battery has to be charged.
In order to prevent a damage of the battery, the instrument switches automatically
off before the battery is completely empty.
8.2 Battery status indication on the display
Battery is empty. The instrument switches automatically off.
Battery is soon empty. The instrument should be charged.
Battery is full.
Instrument is connected with the battery charger and the charging status is
shown on the battery charger.
8.3 Charging
To charge the battery, plug in the charger
plug on the ZRM 6006 and connect the
charger to a wall socket (100 - 240 V, 50 - 60
Hz).
The special plug on the battery cable has a
reverse polarity protection. The ZRM 6006
can be turned off or on while charging.
The battery charger may be connected to the
ZRM 6006 also after finishing the charging
cycle as long as it is still connected to a wall socket. For disconnecting the charger
press on the “push button” on the port for battery charger
8.3.1 Charging status indication / charging cycle
The charge indicator light on the charger shows the charge of the battery in the
charging cycle:
•
Is the indicator light red, the unit is fast charging. The battery charger is in
constant current mode and the charge current is maximum.
•
Is the indicator light orange, the charging cycle “final charge” is active. The
battery is normally approx. 80% charged when the LED-indicator changes to
orange. The battery charger is in constant voltage mode.
•
Is the indicator light green, the battery is fully loaded.
Содержание Zehntner ZRM 6006
Страница 1: ...ZRM 6006 Retroreflectometer RL Qd Instruction Manual from Firmware v 1 3 ...
Страница 2: ......
Страница 15: ...15 ...