5.4. Basic Measurement – Waveform Display
Select “Basic
Measurements”
Start the measurement
Select “Waveform”
Adjust as required
(See below)
Cancel to return to
standard display
Waveform Controls
Y < >
Vertical zoom
X < >
Horizontal zoom
t - +
Manual trigger adjustment
Stop/Start
Note 1: Rx gain may need reducing for vertical zoom to have a noticeable effect.
Note 2: Once adjusted, trigger point does not reset automatically unless a new measurement is started
in the menu “Start the measurement”
Note 3: The waveform is not saved, only the transmission time.
5.5. Basic Measurement – Path Length
If the pulse velocity of the material under test is known, it is possible to measure the path length
between the transducers. The procedure is exactly the same as for measuring pulse velocity ex-
cept that in this case the known pulse velocity is entered.
Select
“Basic Measurements”
Select parameter setting
“pulse velocity”
Enter the pulse velocity
Start the measurement
The display shows:
- Transmission time
- Measured path length
- Received signal level
Stop the measurement
(Not necessary in burst mode)
Save the result or:
Adjust the gain settings.
(See Chapter 3) or:
Start a new measurement