© 2014 Proceq SA
15
4.3.2 Calibration on known Material / Determinate the US velocity
To determine the velocity of an US wave in a material with a known thickness, use the “Calibrating
on” function in the setup screen.
Enter the thickness of the sample with “Edit” and start the calibration process by pressing “Run”.
The sample thickness should be within 2.0 - 80.0mm (0.078 - 3.15 inch).
Once the data acquisition is finished, the velocity and material can be saved and added to the
database of the instrument (see 4.3.1.).
4.3.3 Overview of Different Setup-Items
Calibration On
4
All Modes
Designed for determining the velocity
of an US wave in a material with a
known thickness
2.0 - 80.0 mm
(0.078 - 3.15 inch)
Monitor
5
All Modes Set response conditions for color,
sound or vibration alarm
• Inside: results are within the
set interval
• Outside: results are not in
the set interval
• Off: monitor is switched off
Limit: beginning
6
All Modes
Set the lower limit of the “monitor”
0 - 150 mm (0 - 5.9 inch)
Limit: end
7
All Modes Set the upper limit of the “monitor”
1 - 300 mm (0 - 11.8 inch)
Clear Memory
14
Memory
Erasing the measurement results
Scan beginning
15
A-Scan
Set the beginning of the reflection
area on the screen
0 - 150 mm (0-5.9 inch)
Scan end
16
A-Scan
Set the end of the reflection area on
the screen
5 - 300 mm (0.2 - 11.8 inch)
Gate: beginning
17
A-Scan
Set lower limit
0 - 150 mm (0 -5.9 inch)
Gate: end
18
A-Scan
Set upper limit
1 - 300 mm (0 - 11.8 inch)
Gain
19
A-Scan
Set the amplification of the instrument
inlet path
0 - 80 dB
Type of A-Scan
20
A-Scan
Selection of the signal reflection type
• Fill: reflected in the filled
type
• Outline: reflected in form of
an outline