FACTS ABOUT PRISMA DI-5

TOLERANCE

Prisma DI-5 has a tolerance of max. 1% in the range of +/- 1,000 mm

and max 2% in the range of 1,000-2,000 mm.

HOW TO ACHIEVE OPTIMUM RESULTS

An accurate measurement result can be achieved if both transducer and the measured 

object (Crankshaft) are having the same temperature so that the transducer temperature 

is not changed during the measurement.

If the transducer’s temperature differs from the measured object (Crankshaft)

temperature, an error value of approximately 0.002mm/°C can be experienced.

That’s why the transducer should as far as possible, be temperature-acclimatized to 

the measured object (Crankshaft) temperature before starting with the measurement.

SAFETY INSTRUCTIONS

• Be careful with tools that may cause short circuit at the battery charge port.

• Make sure to use only the original charger supplied with the Prisma DI-5.

TROUBLESHOOTING

• If the display is lighting up without any readings then please press on 

  the ”POWER” button for 3 seconds to switch off the device and try to start it again.

• If the Prisma DI-5 is not used for a long time or not being charged for at least once a 

  year then it may be difficult to start up the device and accordingly a battery reset can  

  probably help to have the Prisma DI-5 up and running again.

  To reset the battery please do as follow

  1.  Remove the left cover of the display unit. 

    The cover which includes the battery charge port.

  2.  Pull out the battery contact and put it back again, the display will 

    start up and go down again.

  3.  Remount the left cover 

Содержание DI-5 Series

Страница 1: ...M A N U A L I N S T R U M E N T DI 5...

Страница 2: ...Battery Charge Port Transducer port Transducer Tip Locking Ring Measuring Tip Extension Bar 10 mm Extension Bar 20 mm Extension Bar 40 mm Extension Bar 80 mm Extension Bar 160 mm INSTRUMENT OVERVIEW...

Страница 3: ...on 6 Fit the transducer between the webs in the punch marks Adjust the screw ring to a reading close to zero and lock it with the locking ring 7 Push ZERO button 8 Rotate crankshaft and check deflecti...

Страница 4: ...arting with the measurement SAFETY INSTRUCTIONS Be careful with tools that may cause short circuit at the battery charge port Make sure to use only the original charger supplied with the Prisma DI 5 T...

Страница 5: ...er 5 if mounted 6 Unscrew the tip 6 from the core 7 and mount the new transducer tip 7 Remount in opposite order 2 3 5 6 7 1 4 MAINTENANCE The calibration certificate supplied with the Prisma DI 5 is...

Страница 6: ...Transducer W min 100 mm including measuring tip 10 mm 434 4758 Transducer Tip 17 mm 458 5000 Extension Bar Set 2x160 mm 1x80 40 20 10 mm 10 mm measuring tip 458 5107 Extension Bar 10 mm 458 5160 Exte...

Страница 7: ...and Recycling of Prisma DI 5 The metal parts to be recycled as metals The electronic parts to be recycled as electronics The plastic case bag to be recycled as plastic The Lithium Ion battery to be r...

Страница 8: ...ner is in the cylinder top Using the Prisma DI Ovality Kit together with Prisma DI 5C and kit you do have an outstanding funktion to load all your measurements into the PC software and print out graph...

Отзывы: