Chapter 6
Advanced Topics
93
93
Shutter expander:
Shutter Mode:
Normal
Trigger expander:
Trigger Response:
No Response, Readout Per Trigger, or Shift Per
Trigger
Summary
Kinetics mode is a powerful feature that allows a burst of subframes to be captured with
microsecond time resolution. However, careful attention must be paid to the optical and
timing considerations, namely
The rows furthest from the serial register need to be illuminated.
The rest of the active area must be optically masked.
Single trigger {Readout Per Trigger} mode allows the capture of a burst of frame
with just one trigger.
In multiple trigger {Shift Per Trigger} mode, the camera requires a trigger for
each "exposure-shift" cycle.
The highest time resolution between kinetic frames is determined by the window
(subframe) size.
Acquisition must be in Safe Mode if you are using WinView or WinSpec.
Binning and ROI selections are supported as in the standard operation.
Spectra-Kinetics (Option)
Spectra-kinetics (standard for the ProEM+:512BK) is a purchasable software option for
ProEM+:512B and 1024B
cameras. When this option is available you will be able to
select Spectra-Kinetics as the readout mode.
ProEM+:512B/BK
ProEM+:1024B
Rows
illuminated
Total number of
kinetic frames
(kinetics)
Total number of
kinetic frames
(spectra-kinetics)
Total number of
kinetic frames
(kinetics)
Total number of
kinetic frames
(spectra-kinetics)
32
32
528
64
1037
64
16
528
32
1037
128
8
528
16
1037
256
4
528
8
1037
Table 4. Comparison of Kinetics and Spectra-Kinetics for ProEM+:512B/BK and
ProEM+:1024B
Unlike standard kinetics mode where binning is performed in the serial register, spectra-
kinetics hardware bins the exposed rows into a single row as the rows are being shifted
into the frame transfer area (or under an external mask). By using this technique, the
spectra-kinetics option is able to capture longer-duration events and provide higher
sensitivity than standard kinetics mode. Use of spectra-kinetics allows the ProEM+
camera to acquire the full masked height’s worth of spectra regardless of the height of the
Содержание ProEM+ EMCCD
Страница 1: ...EMCCD Cameras For Imaging and Spectroscopy ProEM System Manual...
Страница 2: ...4411 0144 Version 1 B January 07 2012 4411 0144...
Страница 11: ...x ProEM System Manual Version 1 B This page intentionally left blank...
Страница 17: ...16 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 77: ...76 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 85: ...84 ProEM System Manual Version 1 B Figure 46 Flowcharts of Safe and Fast Mode Operation...
Страница 99: ...98 ProEM System Manual Version 1 B Figure 63 WinX Vertical Shift...
Страница 121: ...120 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 125: ...124 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 127: ...126 ProEM System Manual Version 1 B ProEM 512BK Kinetics Figure 84 ProEM 512BK Outline Drawing...
Страница 129: ...128 ProEM System Manual Version 1 B ProEM Power Supply Figure 86 ProEM Power Supply Outline Drawing...
Страница 130: ...Appendix B Outline Drawings 129 129 CoolCUBEII optional Figure 87 CoolCUBEII Outline Drawing...
Страница 131: ...130 ProEM System Manual Version 1 B This page intentionally left blank...