Chapter 5
Operation
57
57
will be active. The
Status
bar (at the bottom of the window) displays an icon for
temperature status which reports the current temperature and whether the set
temperature has been reached. Clicking on the icon opens the
Sensor
expander
where the set temperature can be changed.
3.
On the
Shutter
expander, select
Shutter Mode:
Normal
.
4.
Open the
Spectrometer
expander, select the appropriate grating. In this case,
the 300g/mm (Blaze: 750) grating was selected and the center wavelength was
set to 500 nm for a mercury lamp. Use 0.0 nm if using a broadband source.
Confirming the Setup
1.
Turn on the light source at the spectrograph entrance slit.
2.
Click on
Run
to begin previewing the data. Depending on the display settings,
you should see either a spectral band (image) or a graph. Background noise will
decrease as the camera cools to its default temperature.
3.
Turn off the light source. The data display should change to a background noise
pattern or low intensity graph. If this occurs you have confirmed that light
entering the spectrograph is being seen by the camera. Skip Steps 4-5 and
continue to the
“Rotational Alignment and Focusing”
topic on page 58.
4.
If there is little or no difference between the data displayed when the light source
is on or off:
a. Verify that the camera’s manual shutter is open (i.e., the shutter knob should
be pushed in).
b. Verify that the light source has power and is turned on.
c. Verify that the entrance slit is open at least 10 µm.
d. Check the
Exposure Time
(
Common Acquisition Settings
expander
)
.
e. Confirm that
Shutter Mode
is set to
Normal
(
Shutter
expander
)
.
f. Check the shutter cable connections.
g. Verify shutter operation. You should hear the shutter open and close while
Run
is active.
If you hear a shutter operating and you have done Steps a-f, turn the light
source on, wait a minute and then turn the light off while you view the
data display. If the problem is fixed, stop acquisition or continue to the
“Rotational Alignment and Focusing,”
topic on 58.
If you do not hear a shutter operating and you have done Steps a-f,
stop
data acquisition
and continue to Step h.
h. Make sure the spectrograph has an entrance slit shutter. An externally
mounted shutter is easily confirmed. Verifying an internally mounted shutter
requires access to the inside of the spectrograph: refer to the spectrograph
manual for instructions.
5.
If you need more help, contact Customer Support. See page 146 for contact
information.
Содержание ProEM+ EMCCD
Страница 1: ...EMCCD Cameras For Imaging and Spectroscopy ProEM System Manual...
Страница 2: ...4411 0144 Version 1 B January 07 2012 4411 0144...
Страница 11: ...x ProEM System Manual Version 1 B This page intentionally left blank...
Страница 17: ...16 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 77: ...76 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 85: ...84 ProEM System Manual Version 1 B Figure 46 Flowcharts of Safe and Fast Mode Operation...
Страница 99: ...98 ProEM System Manual Version 1 B Figure 63 WinX Vertical Shift...
Страница 121: ...120 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 125: ...124 ProEM System Manual Version 1 B This page intentionally left blank...
Страница 127: ...126 ProEM System Manual Version 1 B ProEM 512BK Kinetics Figure 84 ProEM 512BK Outline Drawing...
Страница 129: ...128 ProEM System Manual Version 1 B ProEM Power Supply Figure 86 ProEM Power Supply Outline Drawing...
Страница 130: ...Appendix B Outline Drawings 129 129 CoolCUBEII optional Figure 87 CoolCUBEII Outline Drawing...
Страница 131: ...130 ProEM System Manual Version 1 B This page intentionally left blank...