![Primes MicroSpotMonitor Plus HighBrilliance Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/primes/microspotmonitor-plus-highbrilliance/microspotmonitor-plus-highbrilliance_original-instructions-manual_1601728061.webp)
61
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
Option
Explanation
Averaging *
If this option is enabled, different algorithms for averaging a plane
measurement can be selected in the drop-down list below. Averaging
over several measurements can be useful, for example, when measuring
a laser with significant power fluctuations.
1. Set the check mark to enable the option
2. Select an algorithm from the drop-down list:
•
Arithmetic mean
: The measured values for each pixel are added
together and divided by the number of planes.
•
Max. intensity per pixel
: The values from all the measurements are
compared for each pixel and only the maximum value for each one
is displayed.
•
Max. lines
: The values from all the measurements are compared for
each line (meaning the line issuing from the measuring device in the
x-direction for example) and only the maximum value for each one is
displayed.
Number of averaged planes *
X
With the
Averaging
option enabled, enter the number of plane
measurements for averaging.
Trigger
Drop-down area for displaying the options of the trigger operation.
The trigger operation is suitable for measuring special pulsed beams.
Detailed information on this topic can be found in the chapter 16.3 on
Drop-down list for trigger selection
This drop-down list is only visible if the drop-down area for displaying the
options of the trigger operation is open.
You can select from the following options:
•
UseNoTrigger
: Switches to the default cw / quasi-cw operation.
•
UsePhotoTrigger
: Switches to a trigger operation in which the
measurement is triggered internally. The signals come from the
photodiode built into the device.
•
UseExternalTrigger
: Switches to a trigger operation in which the
measurement is triggered externally. The trigger signals enter the
device via the BNC socket.
TriggerDelay in ms *
This drop-down list is only visible if the drop-down area for displaying the
options of the trigger operation is open.
The trigger delay specifies the period between the trigger signal and the
beginning of the integration time. Use one of the following options to set
the trigger delay time:
X
Enter a value in the input field (0.012 ms to 400 ms).
X
Use the slider below the input field.
TriggerLevel *
This drop-down list is only visible if the drop-down area for displaying the
options of the trigger operation is open.
The trigger level is the threshold value for triggering a signal via the inter-
nal photodiode. Use one of the following options to set the trigger level:
X
Enter a value in the input field.
X
Use the slider below the input field.
You can determine the threshold value suitable for a measurement task
as follows:
1. Set the trigger level to the maximum value.
2. Decrease the value step by step until the device receives some
trigger signals (lower trigger level).
3. Increase the value until the device does not receive any trigger
signals (upper trigger level).
4. Calculate the trigger level as the arithmetic mean of the two limit
values.
Tab. 9.4:
Options in the
Device control > Advanced
tab of the
Single planes
measuring mode
Содержание MicroSpotMonitor Plus HighBrilliance
Страница 2: ......
Страница 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Страница 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Страница 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Страница 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Страница 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Страница 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Страница 101: ...101 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 with fibre bridge continued B View A...
Страница 102: ...102 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB with fibre bridge continued View B...
Страница 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Страница 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...
Страница 105: ...105 Revision 00 EN 02 2022 MSM HB 35 0 6 15 4 Dimensions of the MSM HB20 with fibre bridge at zmax D...
Страница 106: ...106 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 with fibre bridge continued View D...
Страница 116: ...116 Revision 00 EN 02 2022 MSM HB 35 0 6...