8.5
Safety Circuit PLM In (option)
This connection is only used if you are using a PowerLossMonitor PLM for power measurement.
NOTICE
Danger of damage due to overheating
If the safety circuit is not connected, the device can be damaged due to overheating.
X
Connect the Safety Interlock of the PowerLossMonitor PLM to the MSM+ HB.
X
Check that the laser is switched off properly by the external safety circuit in the event of a
fault.
The Safety Interlock protects from damage by switching off the laser in the following cases:
• The water flow is too low.
Make sure that the Safety Interlock MSM connector on the PowerLossMonitor PLM is connected to the
Safety Circuit PLM-In connector on the MSM+ HB. Use the enclosed 8-pin cable, which is pre-assembled
on both sides.
The interlock connection provides two redundant safety circuits.
8.6
Safety Circuit Fiber Out (option)
This connection is only used if you are using a fiber bridge to connect a laser.
The Safety Interlock protects against damage by switching off the laser in the following cases:
• The fiber is not mounted correctly.
Make sure that the safety interlock connection on the fiber holder is connected to the Safety Circuit Fiber Out
connector at the MSM+ HB. Use the enclosed 3-pin cable.
8.7
Inlet external trigger (In)
BNC connector (view: connector side)
1
2
Pin
Function
1
+5 V (Trigger signal)
2
GND
Tab. 8.3:
Connection socket inlet for an external trigger in the connection panel
8.8
Outlet internal trigger (Out)
BNC connector (view: connector side)
1
2
Pin
Function
1
+5 V (Trigger signal)
2
GND
Tab. 8.4:
Connection socket outlet for the internal trigger in the connection panel
Содержание MicroSpotMonitor Plus HighBrilliance
Страница 2: ......
Страница 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Страница 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Страница 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Страница 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Страница 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Страница 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Страница 101: ...101 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 with fibre bridge continued B View A...
Страница 102: ...102 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB with fibre bridge continued View B...
Страница 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Страница 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...
Страница 105: ...105 Revision 00 EN 02 2022 MSM HB 35 0 6 15 4 Dimensions of the MSM HB20 with fibre bridge at zmax D...
Страница 106: ...106 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 with fibre bridge continued View D...
Страница 116: ...116 Revision 00 EN 02 2022 MSM HB 35 0 6...