35,0(6
Cube L1
25
Revision 03 EN - 03/2021
11.2
Warning message
Warning message
Reason
Safety interlock open
The absorber temperature has exceeded 100 °C. In order to protect
the device, the safety Interlock is switched to the impermissible
operating status.
Tab. 11.2: Warning message
This warning message does not imply an error. The message is automatically
reset as soon as the absorber temperature is below 100 °C again. The safety
interlock is then also automatically switched back to the permissible operating
status.
11.3
Charge status of the lithium-ion battery
The charge status of the battery is displayed in percent. The accuracy of this display
depends on various factors (e.g. temperature, age of the battery, etc.). Recharge the
battery when the charge level reaches 20 percent.
The battery can be charged with a max. charging current of 1.3 A. The battery has a
capacity of 1 000 mAh. The charging time of the battery on the PC via USB connection
with a charging current of 500 mA is approx. 2 hours.
At 100 % charge status, the possible operating time is approx. 6 hours (equivalent to
approx. 100 measurements). When using all power saving functions (see chapter 14.4
on page 39) approx. 15 hours.
Содержание Cube L1
Страница 1: ...Revision 03 EN 03 2021 35 0 6 Cube L1 LaserDiagnosticsSoftware LDS Cube App Original Instructions ...
Страница 2: ......
Страница 3: ...35 0 6 Cube L1 3 Revision 03 EN 03 2021 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE ...
Страница 51: ...35 0 6 Cube L1 51 Revision 03 EN 03 2021 18 Declaration of conformity ...
Страница 55: ...35 0 6 Cube L1 55 Revision 03 EN 03 2021 ...
Страница 56: ...PRIMES GmbH Max Planck Str 2 64319 Pfungstadt Germany Tel 49 6157 9878 0 info primes de www primes de ...