![Primes Cube L1 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/primes/cube-l1/cube-l1_original-instructions-manual_1601735021.webp)
35,0(6
Cube L1
21
Revision 03 EN - 03/2021
Please observe the information on minimum and maximum energy input per measure-
ment according to chapter 12.2.3 on page 30.
A suitable 2 m long connection cable is included.
NOTICE
Damaging/Destruction of the device
If the safety interlock is not connected, the device can be damaged or
destroyed due to overheating.
X
When connecting the laser control, please ensure that the laser is turned
off in case of an interruption of the connection.
Pin diagram safety interlock
plug (view connector side)
Pin
Wire color
Function
1
4
3
4
Black
Mutual pin
1
Brown
Connected with pin 4, when ready for
operation
3
Blue
Connected with pin 4, when in safety
interlock mode (absorber too hot)
Tab. 9.1: Pin assignment of the safety interlock plug
9.2
Micro-USB socket
You can charge the battery of the device on the PC via the micro USB socket. Only use
the supplied USB cable to charge the battery.
When using the optional LaserDiagnosticsSoftware LDS (not included in delivery), the
device communicates with the LDS via the micro-USB socket or Bluetooth.
You will find the PRIMES USB-driver for all USB-capable devices on the PRIMES web-
site at: https://www.primes.de/en/support/downloads/software.html.
Содержание Cube L1
Страница 1: ...Revision 03 EN 03 2021 35 0 6 Cube L1 LaserDiagnosticsSoftware LDS Cube App Original Instructions ...
Страница 2: ......
Страница 3: ...35 0 6 Cube L1 3 Revision 03 EN 03 2021 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE ...
Страница 51: ...35 0 6 Cube L1 51 Revision 03 EN 03 2021 18 Declaration of conformity ...
Страница 55: ...35 0 6 Cube L1 55 Revision 03 EN 03 2021 ...
Страница 56: ...PRIMES GmbH Max Planck Str 2 64319 Pfungstadt Germany Tel 49 6157 9878 0 info primes de www primes de ...