20
Using RFID
Note:
Write success is less likely as you move the marker away
from the center of the label.
Important Note:
The remaining items are optional and only
necessary if encoding problems are present. If you are loading a new
roll with the exact same specifications as the last roll, none of these
steps will be necessary.
Calibrate.
Open the PTStatus Monitor. Go to Start - All
Programs - Primera Technology - PTStatus. Click on the
RFID tab.
Click "Calibrate". PTStatus will automatically calibrate optimal
vertical position and Antennae Power. This process may take
up to three minutes. During this time the RFID read/write
offset position and antennae power are automatically adjusted.
After this, ten write and ten verification reads are performed
on the same label.
Writing Test.
If further testing is desired or if the automatic
calibration failed you can perform a writing test. Select the
Writing Test button to start.
The RX900 will write to the same label continuously until "Stop
Test" is selected. The results of each test will be displayed in
PTStatus. During this time you can adjust the Slide Bar if
necessary until you receive a successful test. If the test is not
successful you will see a "Write Failed" error.
Set Marker
approximately
1/8" (3mm) left
or right from
center chip of
antennae.