GB 04335 - Edition 02 - March 12
1/5
C 500
Factory configuration of the leak
detector parameters
The following list indicates the factory configuration of the leak
detector parameters.
When the leak detector is switched off, all set parameters are
memorized and values are kept for the next start-up.
We advice you to note in the “Customer modification“ column, the
parameter values modified for your application.
The operator can save and load different leak detection configurations.
For that, use the functions "Save LD parameters" in the "SD card"
advanced menu (
C 406
)
.
The saved values are the set values at the saving moment.
Parameters
configuration
Parameters
Configuration
Factory
Vacuum reject point
1.0E-07
Sniffing reject point
1.0E-04
Probe clogged reject point
1.1E-06
Alarm
3 - On
Digital voice
4 - On
He Max.
Off
He Max. reject point
1.0E-05
C 401
adixen Vacuum Products - ASM 380 Operating instructions