Advanced Menu
GB 04338 - Edition 03 - March 12
s¬
It is advised to perform an autocalibration:
- at the detector switching on in order to make sure that it is in
correct conditions use
- to optimize the measure reliability for high sensitivity tests
- in case of doubt about the good leak detector functioning (ability
to detect correctly a He leak). At any time, an internal calibration
can be launched.
- in case of continuous and intensive use: launch an internal calibra-
tion at each team beginning (8 hours of work).
Other advanced methods
It is also possible to do a detector autocalibration with an external
calibrated leak
E 410
.
Calibration = manual
Calibration is made manualy,
OPERATION RESERVED TO SERVICE CENTERS AND EXPERTS ONLY.
The calibration checking is deactived.