GB 04267 - Edition 02 - July 2011
7/7
B 400
Starting-up / Switching off of the leak
detector
Your leak detector is multivoltage (from 90 to 240 V ± 10 %):
connect it to the power plug.
During the first starting up after reception, you can adjust the unit,
the date, the hour and the language of the displays and the digital
voice.
These parameters can afterward be modified if necessary
(
C 303
).
After an unused storage period, there is additional time at the start-
up for outgassing.
The countdown before using the leak detector is displayed.
The operator can cancel this additional time by pressing
.
But it is not recommended.
It is not abnormal that after an unused storage period, the
calibration fails. In this case, wait 30 minutes and do it again.
If you have never used a leak detector, know that there is no risk
of starting your detector and discovering the menus even without
particular knowledge. But we advise to read chapters B and C.
If you have already used this type of product, do not hesitate to
discover the menus by yourself. Read the chapter C.
The leak detector can be switched off at any time by depressing the
main power switch to 0 (OFF) but it is advised (especially when the
leak detector may not be used for an extended period of time) to
stop it following the procedure described below.
Place the blank-off flange on the inlet port or link the hose
holder to the detector inlet port (
B 240
)
.
Start a test cycle.
Wait until the leak detector reaches the more sensitive test mode
Make sure that the inlet vent is OFF.
Stop the test cycle.
Stop the leak detector.
Wait before moving the detector so as to make sure of the
complete secondary pump stop.
The leak detector is so under vacuum avoiding an important
degassing time in a future use.
Starting up
1
st
starting up
Starting up after an
unused/
storage period
Getting started with the
leak detector
Switching off
recommanded
adixen Vacuum Products - ASM 380 Operating instructions