GB 04493 - Edition 03 - March 12
11/15
B 303
37 pin Input/Output interface
Possible values for each input:
Value
Function
None
HV test
Start/Stop a test in hard vacuum mode
Calibration
Start an autocalibration
Zero
Activate/Deactivate Zero function
He Memo
Activate/Deactivate Memo function
Dynamic cal.
Start a dynamic calibration
(2)
Sniffer test
Start/Stop a test in sniffing mode
Filament
Force stop/start filament
GL mode
Force leak detector
(1)
in Gross Leak test mode
NR mode
Force leak detector
(1)
in Normal test mode
HS mode
Force leak detector
(1)
in High Sensitivity test mode
Inlet vent
Opening/closing of the inlet vent valve
(1)
Rec. Graph
Start/stop data recording
Save Graph
Save recorded data on a SD card
(1) ASI 30 not concerned
(2) Only ASI 30 concerned
4 adjustable digital
transistor outputs.
Digital Output
5 adjustable digital relay
outputs, dry contact type.
adixen Vacuum Products - ASM 380 Operating instructions