Warning: ERROR message indicates that the accessories are not functioning
properly (cable breakage) or the contact area is dirty.
Warning:
• Wipe the surface of the device and the measuring cable with a clean cloth,
lightly moistened with a non-aggressive disinfectant.
• Do not apply any liquid or spray directly on the device, especially on the
display.
• Do not use high-proof alcohol for disinfection.
Warning:
• The measuring cable cannot be autoclaved.
• Use of agents other than specified above may cause damage to the equipment
and its accessories.
• Do not use heat, radiation, formaldehyde or ethylene oxide or plasma
sterilization methods.
Fig. 34 Accessories ERROR message
Fig. 33 Accessories are OK
For assistance call your local distributor or
manufacturer directly.
CHECK mode will exit after 3s automatically.
Cables CHECK will start automatically and the results – either OK (Fig. 33) or ERROR (Fig. 34) -
will be shown on the screen.
OK
ERROR
4 MAINTENANCE, CLEANING AND STERILIZATION
4.1 General
Apica Pro is maintenance free and does not contain user serviceable parts.
Measurement cable and the surface of the device should be cleaned using tissue or soft cloth
soaked with aldehyde free disinfecting and detergent solution (bactericidal and fungicidal).
Remove the lip clip, file clip from the measurement cable prior to the reprocessing. The accessories
lip clip and file clip must be cleaned, disinfected and sterilized prior to each use. This also applies
to the first use of accessories. Thorough cleaning and disinfection are essential prerequisites for
effective Sterilization. The specific instructions for cleaning/ Sterilization must be applied according
to the instructions in chapter “Cleaning, Disinfection and Sterilization” 5.2 In addition, the
operating instructions of the devices used in your practice must be followed.
As part of your responsibility for accessories sterility, always ensure that only validated methods for
cleaning/disinfection and Sterilization are used, that devices (disinfector, sterilizer) are regularly
serviced and inspected and that the validated parameters are maintained with each cycle.
In addition, always observe the validated legal regulations and regulations on hygiene relating to
your practice or the hospital. This applies in particular to the guidelines regarding effective prion
inactivation.
For your own safety, always wear protective gloves, glasses, mask when handling contaminated
accessories.
Note: Service and repair should be provided by factory trained service
personnel only.