Helios Large Display Strain Gauge Meter Instruction Manual
PD2-6100
24
Display Functions & Messages
The meter displays various functions and messages during setup, programming, and operation. The
following table shows the main menu functions and messages in the order they appear in the menu.
Display
Parameter
Action/Setting Description
setup
Setup
Enter
Setup
menu
Input
Input
Enter
Input
selection menu
Polar
Polar
Enter unipolar or bipolar selection menu
uni
Unipolar
Press Enter to select operation with positive signals only (0-30 mV)
bi
Bipolar
Press Enter to select operation with positive & negative signals (±15
mV)
range
Range
Enter range selection menu
15 mv
15 mV
Set meter for 15 mV input (uni/bi)
25 mv
25 mV
Set meter for ±25 mV input (bi)
30 mv
30 mV
Set meter for 30 mV input (uni)
150mv
150 mV
Set meter for 150 mV input (uni/bi)
250mv
250 mV
Set meter for ±250 mV input (bi)
300mv
300 mV
Set meter for 300 mV input (uni)
ratio
Ratiometric
Press Enter to select ratiometric operation to compensate for excitation
changes.
d-SCAL
Dual-scale
Press Enter to select dual-scale display for some level applications
(Select Yes or No)
units
Units
Select the display units/tags
Dec pt
Decimal point
Set decimal point
Pv 1
PV1
PV1 decimal point (Level)
Pv 2
PV2
PV2 decimal point (Level)
prog
Program
Enter the
Program
menu
SCALE
Scale
Enter the
Scale
menu
SCAL 1
Scale 1
Enter the
Scale
menu for PV1
SCAL 2
Scale 2
Enter the
Scale
menu for PV2
Cal
Calibrate
Enter the
Calibration
menu
Inp 1
Input 1
Calibrate input 1 signal or program input 1 value
Dis 1
Display 1
Program display 1 value
Inp 2
Input 2
Calibrate input 2 signal or program input 2 value (up to 32 points)
Dis 2
Display 2
Program display 2 value (up to 32 points)
Error
Error
Error, calibration not successful, check signal or programmed value
dsplay
Display
Enter the
Display
menu
Line 1
Display Line 1
Press Enter to assign the display line 1 parameter (default: PV)
Line 2
Display Line 2
Press Enter to assign the display line 2 parameter (default: engineering
units)
d-Inty
Display intensity
Set display intensity level from 1 to 8
RELaY
Relay
Enter the
Relay
menu
RLY 1
Relay 1
Relay 1 setup
Act 1
Action 1
Set relay 1 action
Auto
Automatic
Set relay for automatic reset
A-nman
Auto-manual
Set relay for automatic & manual reset any time
LatCH
Latching
Set relay for latching operation
Lt-CLr
Latching-cleared
Set relay for latching operation with manual reset only after alarm
condition has cleared