PD2-6100
Large Display Strain Gauge Meter Instruction Manual
17
Installation
Figure 18 shows the connection of a meter to a PC using the RS485 output connector and a PDA7485
RS-232 to RS-422/485 converter in an RS-422 network. Figure 19 shows the connection of several
meters to a PC using a PDA7485 RS-232 to RS-422/485 converter in an RS-485 network.
When using more than one meter in a multi-drop mode, each meter must be provided with its own unique
address. The meter address (Slave ID) may be programmed between 1 and 247. The transmit delay may
be set between 0 and 199 ms. The parity can be set to even, odd, or none with 1 or 2 stop bits.
To change the meter address:
1. Press and hold the
Menu
button for three seconds to access
Advanced Features
menu of the meter.
2. Press
Up
arrow until Serial (
SEriaL
) menu is displayed and press
Enter
,
Addres
is displayed.
3. Press
Enter
to change meter address using
Right
and
Up
arrow buttons. Press
Enter
to accept.
4. Press
Menu
button to exit and return to Run Mode.
Figure 18. RS-422 or RS-485 Wiring
Notes:
1. Termination resistors are optional and values depend on the cable length and characteristic
impedance. Consult the cable manufacturer for recommendations.
2. Refer to RS-232 to RS-422/485 Converter documentation for further details.
3. Use shielded cable, twisted-pairs plus ground. Connect ground shield only at one location.