Helios Large Display Dual-Input Meter Instruction Manual
PD2-6060
52
Round Horizontal Tank Linearization (
rHt
)
This function is used to calculate volume in a round horizontal tank with flat ends. The volume is
calculated based on the diameter and length of the tank. The tank’s dimensions can be entered in inches
or centimeters; the meter automatically calculates the volume in gallons or liters. After entering the
dimensions, complete the scaling process with the display values calculated by the meter. The meter can
be re-scaled to display the volume in any engineering unit without the need to re-enter the dimensions
again.
Note: After Scale is displayed continue pressing the Enter button until the meter completes the scaling of
the input and display values.
Changing the Volume from Gallons to Liters
In the above graphic, entering the 48" for the diameter and 120" for the length of the round horizontal
tank, the meter automatically calculates that the volume of the tank is 940.02 gallons.
1. Convert gallons to liters
2. 1 US gallon = 3.7854 L
3. 940.02 gal = 3558.4 L
4. Go to the
Setup
menu and change the decimal point to 1 decimal.
5. Go to the
Program
–
Scale
menu and press Enter until
dis 2
is shown on the main display.
6. Press Enter and change the display 2 value to 3558.4.
7. The meter is now displaying the volume in liters.
Note: The display can be scaled to display the volume in any engineering units.
www.
GlobalTestSupply
.com
Find Quality Products Online at: