8 Data transfer to peripheral devices
62
8.3
Remote control-commands
Command
Function
ACKn
Handshake n=0 off; n = 1 on
CAL
Start calibration (only where INT or EXT selected)
DN
Reset weight display
D..........
Describe weight display (right-aligned)
@N
Reset Info display
@.........
Describe Info display
In
Set FLOATINGDISPLAY time n=0
t=0.04 s
n=1
t=0.08 s
n=2
t=0.16 s
n=3
t=0.32 s
N
Reset balance
OFF
Switch off balance
ON
Switch on balance
PCxxxx
Enter anti-theft code
PDT
Print out date and time
PRT
Start printing (Press “Print” key)
PST
Start print status
Pn (ttt.t)
Set print mode
n=0 Individually print each value (unstable)
n=1 Individually print each value (stable)
n=2 Print after change of load
n=3 Print after each integration period
n=4 Print on time basis in s (ttt.t)
R%k
Set current weight=100%
with k=0...7 decimal places (k=A: use automatic
positioning of decimal point)
REF%k rrr...
Set reference weight rrr... for 100%
with k=0...7 decimal places (k=A: use automatic
positioning of decimal point)
Rnnn
Set current weight=nnn items
REFrrr...
Set reference weight rrr... for 1 item