9 Error messages and correction of faults
■
65
Tab. 9.1 Possible faults and their removal
Results of
weighing are
clearly incor-
rect
• The balance was not correctly tared
• The balance is not correctly levelled
• The calibration is no longer correct
• There are strong temperature variations
There is no
display or only
dashes
• The stability control (Balance functions) is set
too sensitively
• The time selected for FLOATINGDISPLAY is
unsatisfactory
Configuration
menu cannot
be changed
• The password lock is activated in the configu-
ration menu
The display
flashes conti-
nuously du-
ring calibration
• The balance location is not quiet enough (Inter-
rupt calibration with «
ON/OFF
» and relocate
the balance in a better position)
• Use of an imprecise calibration weight (only
applies to external calibration)
Fault
Possible cause