26
27
PROGRAMMABLE TRANSMITTER
PRETRANS 5116
• Input for RTD, TC, mV, Ohm, potmeter, mA and V
• 2-wire supply > 16.5 V
• 2.5 V reference voltage
• Output for current, voltage and 2 relays
• Universal AC or DC supply
Application:
• Linearised, electronic temperature measurement with RTD or TC sensor.
• Conversion of linear resistance variation to a standard analogue current / volt-
age signal, i.e. from solenoids and butterfly valves or linear movements with
attached potentio meter.
• Power supply and signal isolator for 2-wire transmitters.
• Process control with 2 pairs of potential-free relay contacts which can be
configured to suit any function.
• Galvanic separation of analogue signals and measurement of floating signals.
Technical characteristics:
• Within a few seconds the user can program PR5116 to suit the specific appli-
cation. As the 5116 is designed with electronic hardware switches, it is not
necessary to open the module for setting of DIP switches.
• A green front LED indicates normal operation and malfunction. A yellow LED
is ON for each active output relay.
• Continuous check of vital stored data for safety reasons.
• 3-port 3.75 kVAC galvanic isolation.
Mounting / installation:
• Mounted vertically or horizontally on a DIN rail. As the modules can be
mounted without any distance between neighbouring units, up to 42 modules
can be mounted per metre.
27
HOW TO DISMANTLE SYSTEM 5000
First, remember to demount the connectors with hazardous voltages.
Picture 1:
Detach the module from the DIN rail
by lifting the bottom lock.
26